IAC Silica Analyzer Series 1400

  • Accurate, Precise and Verifiable Results
  • Full Featured – analog outputs; system and concentration alarms; optional RS422/485 interfaces
  • Fully Automatic and/or Manual Calibration—no specialized operator knowledge necessary
  • Low Reagent Consumption—extremely low cost of operation
  • Photometric Method— well documented, accepted
  • Batch Analysis—more accurate than continuous flow systems
  • No Reagent Disposal—all reagents consumed completely
  • User Adjustable Flow Control
  • Simple, Easy Grab Sample Capability
  • Stand- by mode as a result of sample flow loss
  • Fully Mounted—ready to use right from box
  • 2 Silica Analyzer Configurations—for samples with or without Phosphate—can be operated manually or fully automatic
  • Optional 2-channel system

measurement

Measuring Range: 0–500 ppb
0–5 ppm
Resolution: 0.01 ppb / 0.01ppm
Accuracy: 10% of reading
or +/- 0.5 ppb
Precision: 5% of reading
or +/- 0.2 ppb
Method: on-line colorimetric
Measuring volume: 20 ml
Calibration volume: 1000 ml
Reagent use: 5 liter/month
Number of Reagents: 2
with phosphate
interference: 3

sample conditions

Temperature: 5 … 40 ° C
Pressure: 0.1 … 6 bar (90 psi)
Flow rate: min. 20 ml / min.
Particles: less than 10 ppm,
no oil, no grease
pH- value: between 3 and 8

principle of operation

Detection method
The concentration measurement of reactive silica is based on the well documented color reaction with ammoniummolybdate.

To improve the lower limit of detection, a reducing reagent is added to convert the yellow molybdosilic acid to molybdenum blue.

In the presence of phosphate, an additional reagent is added prior to the reducing agent to remove the phosphate interference.

The analysis is batch-wise, which gives the most stable results compared with other continuous flowing systems.

Specifications are subject to change without notice.

electronics

Display: LCD Display

Bar/numeric mode

Graphic mode

Alarms: 2 potential free relay outputs
one concentration alarm and
one system error
Contact rating: 500 VA, 250 V-2A
Outputs: 4 - 20 mA
free concentration range
Max. load 600 ohms
Resolution 6 μA (12 bit)
Options: TTL or RS422/485
Profibus-DP Interface
Multi-channel

electromagnetic compatibility

Complies with EN 50081-1 and EN 50082-2 for unrestricted use in rural and industrial areas.

dimensions

36" (914mm) H x 24" (610mm) W x 11" (279mm) D